Effects of heterogeneities in dose distributions under nonreference conditions: Monte Carlo simulation vs dose calculation algorithms

dc.TypeArticlept_BR
dc.contributor.authorCarpio, Cristiano Queiroz Melo dos Reis
dc.contributor.authorNicolucci, Patrícia
dc.contributor.authorFortes, Saulo Santos
dc.contributor.authorSilva, Leonardo Peres da
dc.date.accessioned2023-09-06T14:49:19Z
dc.date.available2023-09-06T14:49:19Z
dc.date.issued2019
dc.descriptionp. 74–82.: il. color.
dc.description.abstractThe purpose of this study is to evaluate the performance of dose calculation algorithms used in radiotherapy treatment planning systems (TPSs) in comparison with Monte Carlo (MC) simulations in nonelectronic equilibrium conditions. MC simulations with PENELOPE package were performed for comparison of doses calculated by pencil beam convolution (PBC), analytical anisotropy algorithm (AAA), and Acuros XB TPS algorithms. Relative depth dose curves were calculated in heterogeneous water phantoms with layers of bone (1.8 g/cm3) and lung (0.3 g/cm3) equivalent materials for radiation fields between 1 × 1 cm2 and 10 × 10 cm2. Analysis of relative depth dose curves at the water-bone interface shows that PBC and AAA algorithms present the largest differences to MC calculations (uMC = 0.5%), with maximum differences of up to 4.3% of maximum dose. For the lung-equivalent material and 1 × 1 cm2 field, differences can be up to 24.3% for PBC, 11.5% for AAA, and 7.5% for Acuros. Results show that Acurus presents the best agreement with MC simulation data with equivalent accuracy for modeling radiotherapy dose deposition especially in regions where electronic equilibrium does not hold. For typical (nonsmall) fields used in radiotherapy, AAA and PBC can exhibit reasonable agreement with MC results even in regions of heterogeneities.pt_BR
dc.identifier.citationCARPIO, Cristiano Queiroz Melo dos Reis et al. Effects of heterogeneities in dose distributions under nonreference conditions: Monte Carlo simulation vs dose calculation algorithms. Medical Dosimetry, v. 44, p. 74–82, 2019.
dc.identifier.issn1873-4022
dc.identifier.urihttps://ninho.inca.gov.br/jspui/handle/123456789/14771
dc.publisherMedical Dosimetrypt_BR
dc.subjectDose Calculation Algorithmspt_BR
dc.subjectAlgoritmos de Cálculo de Dosept_BR
dc.subjectHeterogeneity Correctionpt_BR
dc.subjectCorreção de Heterogeneidadept_BR
dc.subjectTreatment Panning Systemspt_BR
dc.subjectSistemas de Tratamento Panorâmicopt_BR
dc.subjectMonte Carlo simulationpt_BR
dc.subjectSimulação de Monte Carlopt_BR
dc.subjectRadiotherapypt_BR
dc.subjectRadioterapiapt_BR
dc.titleEffects of heterogeneities in dose distributions under nonreference conditions: Monte Carlo simulation vs dose calculation algorithmspt_BR

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Effects of heterogeneities in dose distributions under nonreference conditions Monte Carlo simulation vs dose calculation algorithms.pdf
Size:
1.98 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.6 KB
Format:
Item-specific license agreed upon to submission
Description: