Evaluation of gloves as a water bag coupling agent for therapeutic ultrasound

dc.TypeArticlept_BR
dc.contributor.authorLima, Lúcio Salustiano de
dc.contributor.authorOliveira, Débora Paulino
dc.contributor.authorCosta Júnior, José Francisco Silva
dc.contributor.authorPinto, Pâmela Alves
dc.contributor.authorOmena, Thaís Pionório
dc.contributor.authorCosta, Rejane Medeiros
dc.contributor.authorKrüger, Marco Antônio Von
dc.contributor.authorPereira, Wagner Coelho de Albuquerque
dc.date.accessioned2022-05-05T14:22:59Z
dc.date.available2022-05-05T14:22:59Z
dc.date.issued2017
dc.descriptionp. 42-49.: il. color.
dc.description.abstractTherapeutic ultrasound (TUS) is a widespread modality in physiotherapy, and the water bag technique is a coupling method employed in the presence of anatomical irregularities in the treatment area. The aim of the present study is to evaluate the acoustic attenuation of the water bag and its effectiveness as a TUS coupling agent. Methods: The rated output powers (ROPs) of the TUS equipment were evaluated based on IEC 61689. Then, a radiation force balance was used to measure ROP with and without a water bag (latex and nitrile gloves filled with deionized water) between a TUS transducer and the cone-shaped target of the balance. Each experiment was performed five times for each nominal power (0.5, 1.0, 2.0, 3.0, 4.0, 5.0, 6.0, and 7.0 W) and in the following configurations: without the water bag (A), with nitrile gloves and with (B) and without (C) a height controller, and latex gloves with (D) and without (E) height controller. ROPs obtained in different media were compared. Results: The highest relative error of ROP was 16.72% for 0.5 W. Although the power values of the equipment were within the range recommended by IEC, there was a significant difference between the ROP values measured with A and with B, C and D. Conclusion: As intensity differences below 0.5 W/cm2 are considered clinically not relevant, conditions A, B, C, D, or E can be used interchangeably.
dc.identifier.citationLIMA, Lúcio Salustiano de et al. Evaluation of gloves as a water bag coupling agent for therapeutic ultrasound. Res. Biomed. Eng., v. 33, n. 1, p. 42-49, 2017.
dc.identifier.issn2446-4740
dc.identifier.urihttp://sr-vmlxaph03:8080/jspui/handle/123456789/6826
dc.publisherRes. Biomed. Eng.pt_BR
dc.subjectUltrasonicspt_BR
dc.subjectUltrassompt_BR
dc.subjectMethodspt_BR
dc.subjectMétodospt_BR
dc.subjectPotencypt_BR
dc.subjectPotênciapt_BR
dc.subjectAcousticspt_BR
dc.subjectAcústicapt_BR
dc.titleEvaluation of gloves as a water bag coupling agent for therapeutic ultrasoundpt_BR

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Evaluation of gloves as a water bag coupling agent for therapeutic ultrasound..pdf
Size:
1.31 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.6 KB
Format:
Item-specific license agreed upon to submission
Description: