Controle de qualidade em RapidArc com simulador de corpo humano antropomórfico Alderson utilizando filme radiocrômico em comparação ao MATLAB

dc.TypeArticlept_BR
dc.contributor.authorGarcia, Paulo Lazaro
dc.contributor.authorSilva, Leonardo Peres da
dc.contributor.authorAlves, Victor Gabriel Leandro
dc.contributor.authorSantos, Maíra Ribeiro dos
dc.contributor.authorTrindade, Cássia
dc.contributor.authorMartins, Laís Paulino
dc.contributor.authorBatista, Delano Valdivino Santos
dc.date.accessioned2023-01-24T19:11:22Z
dc.date.available2023-01-24T19:11:22Z
dc.date.issued2012
dc.description.abstractEste trabalho apresentou o controle de qualidade para RapidArc utilizando simulador de corpo humano Alderson e filme radiocrômico como uma alternativa de sistema de aprovação final do planejamento de tratamento para tumores de crânio. Deste modo, foram comparadas as distribuições de dose fornecidas pelo sistema de planejamento com aquelas medidas pelo filme radiocrômico. A análise do índice gama (Γ índex) para verificar a aceitabilidade dos cálculos de distribuição de dose, apresentaram 95% de pontos de aprovação, sendo a maioria dos pontos de não conformidade em regiões próximas das bordas do PTV. Podemos tentar justificar tais pontos de não conformidade aos aspectos de transmissão das lâminas devido à região das bordas apresentarem perdas significativas se comparados às áreas centrais. O Programa MATLAB também se mostrou uma ferramenta eficiente para tais medidas e pode ser usado em programas de controle de qualidade.pt_BR
dc.identifier.issn1984-9001
dc.identifier.urihttps://ninho.inca.gov.br/jspui/handle/123456789/12450
dc.publisherRevista Brasileira de Física Médica.2012;6(3):141-4pt_BR
dc.subjectControle de Qualidadept_BR
dc.subjectQuality Controlpt_BR
dc.subjectFilmes Cinematográficospt_BR
dc.subjectMotion Picturespt_BR
dc.subjectRadioterapiapt_BR
dc.subjectRadiotherapypt_BR
dc.subjectsimulador antropomórficopt_BR
dc.subjectRapidArcpt_BR
dc.terms.abstractThis paper presented the quality control for RapidArc using an Alderson human body phantom and radiocromic film as an alternative system to approve the treatment plan for brain tumor. Thus, it was comprised the dose distributions provided by the treatment planning system with those measured by the film radiocromic. The gamma index (Γ) analysis, to verify the acceptability of the dose distribution, was 95% of approved points, with the mostly non-compliance points in regions near the PTV’s edges. These non-compliance points may be associated to transmission blades aspects, because the regions near the edges present significant losses compared to the central areas. Also, MATLAB has proved an effective tool for that measurements and it can be used in quality assurance programs.pt_BR
dc.titleControle de qualidade em RapidArc com simulador de corpo humano antropomórfico Alderson utilizando filme radiocrômico em comparação ao MATLABpt_BR
dc.title.alternativeQuality assurance in RapidArc with Alderson anthropomorphic phantom using radiochromic film in comparison to MATLABpt_BR

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Controle de qualidade em RapidArc com simulador de corpo humano antropomórfico Alderson utilizando filme radiocrômico em comparação ao MATLA.PDF
Size:
2.35 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.6 KB
Format:
Item-specific license agreed upon to submission
Description: