Dosimetric evaluation of electron beam Monte Carlo isodoses distribution based on thermoluminescent dosimetry
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Braz. J. Rad. Sci
Abstract
Electron beams are applied in radiotherapy treatments where superficialized doses are desirable as well as
deeper tissues sparing. Monte Carlo method has been recently implemented in the TPS as the algorithm for
electron beam isodoses calculation at the National Institute of Cancer in Brazil. This work comprises an
additional method of evaluation of the isodoses generated by the TPS, considering the irradiation of a non-flat
surface. A 6-Gy electron beam breast irradiation was planned and delivered to an anthropomorphic phantom
into which 36 capsules with TLD100 were inserted. The TLD calibration curve was performed from 100 cGy to
700 cGy. In an overall analysis, eMC algorithm isodose distribution results agreed with TLD measured doses.
75% of measured data met the 5% accuracy criteria, which can be considered in good agreement with the
recommended uncertainties involved in a radiotherapy treatment. Considered all collected data, 8 TLDs received
extremely low doses and due to tissue attenuation of the electron beam, the TLD measurements might have
additional uncertainties. If these data are not taken into account, 100% of the measured doses, considering the
uncertainties, meet the 5% accuracy criteria. About the 3 methods for isodose smoothing available in the eMC
algorithm, the low level of dose smoothing option provides the best matching with measured data, consisting of
the most reliable option. Considering the AAPM recommendations eMC algorithm offers an accurate solution
for isodoses calculation, within a 5% accuracy.
Description
p. 1-18.: il. p&b. e color.
Citation
GONÇALVES, Flavia dos Reis et al. Dosimetric evaluation of electron beam Monte Carlo isodoses distribution based on thermoluminescent dosimetry. Braz. J. Rad. Sci, p. 1-18, mar. 2020.