Title: A simple method to back-project isocenter dose of radiotherapy treatments using EPID transit dosimetry
Authors: Rosa, Luiz Antonio Ribeiro da
Silveira, Thiago Bernardino da
Cerbaro, Bianca de Quadros
Keywords: Epidemiologia
Epidemiology
Dosimetria
Dosimetry
Sistema Porta
Portal System
Transit Dosimetry
Radioterapia
Radiotherapy
Issue Date: 2017
Publisher: Brazilian Journal of Radiation Sciences
Citation: ROSA, Luiz Antonio Ribeiro da; SILVEIRA, Thiago Bernardino da; CERBARO, Bianca de Quadros. A simple method to back-project isocenter dose of radiotherapy treatments using EPID transit dosimetry. Brazilian Journal of Radiation Sciences, p. 1-18, 2017.
Abstract: The aim of this work was to implement a simple algorithm to evaluate isocenter dose in a phantom using the back-projected transmitted dose acquired using an Electronic Portal Imaging Device (EPID) available in a Varian Trilogy accelerator with two nominal 6 and 10 MV photon beams. This algorithm was developed in MATLAB language, to calibrate EPID measured dose in absolute dose, using a deconvolution process, and to incorporate all scattering and attenuation contributions due to photon interactions with phantom. Modeling process was simplified by using empirical curve adjustments to describe the contribution of scattering and attenuation effects. The implemented algorithm and method were validated employing 19 patient treatment plans with 104 clinical irradiation fields projected on the phantom used. Results for EPID absolute dose calibration by deconvolution have showed percent deviations lower than 1%. Final method validation presented average percent deviations between isocenter doses calculated by back-projection and isocenter doses determined with ionization chamber of 1,86% (SD of 1,00%) and -0,94% (SD of 0,61%) for 6 and 10 MV, respectively. Normalized field by field analysis showed deviations smaller than 2% for 89% of all data for 6 MV beams and 94% for 10 MV beams. It was concluded that the proposed algorithm possesses sufficient accuracy to be used for in vivo dosimetry, being sensitive to detect dose delivery errors bigger than 3-4% for conformal and intensity modulated radiation therapy techniques.
Description: p. 1-18.: il. color. e p&b.
URI: http://sr-vmlxaph03:8080/jspui/handle/123456789/8058
ISSN: 2319-0612
Appears in Collections:Artigos de Periódicos da área de Fisioterapia



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.